| EIS |
EUV Imaging Spectrometer - Fact sheet |
Table 1. EUV Imaging Spectrometer Overview
| Telescope |
| Off-axis parabola |
15 cm diameter; half area:170-210Å, the other half:250-290Å |
| Focal length |
1939mm |
| Off-axis distance |
70 mm |
| Plate scale at slit |
9.37 microns/arcsec |
| Multi-layer coating |
Mo/Si pairs
Peak efficiency at 19.5nm for one half area and 27.0nm for the other |
| Slits & Slots |
1", 2", 40", and 266" width at primary focus |
| Fine mirror scan range |
± 295" solar image motion in east-west direction for raster scan |
| Large FOV translation |
± 890" shift of FOV center in east-west direction |
| Spectrometer |
| Grating |
Toroid surface, holographic and laminar ruling |
| 4200 grooves/mm |
| Wavelength range |
Maximum coverage |
170-210Å, 250-290Å |
| Useful coverage |
180-204Å, 250-290Å |
| Dispersion |
1.65 Å/mm or 0.0223 Å/CCD pixel |
| Pixel equivalent width |
34.3km/s @195Å, 26.1km/s @256Å, and 23.6km/s @284Å |
| Plate scale at CCD |
1.0" / CCD pixel |
| CCD camera |
CCD |
Device |
Marconi 42-20, MPP |
| Format (Spatial × Spectral) |
1024 × 2048 pixel for each CCD |
| Pixel size |
13.5 micron |
| Readout electronics |
Window size |
512 × 2048 max for each CCD from 1024 × 2048 area |
| ICU in spacecraft bus |
Number of spectral windows |
25 maximum |
* FOV of 1024" is guaranteed in north-south direction on the CCD.
Height of 512" out of 1024" can be read in a single exposure.
Table 2. EIS Instrumental Performance
| Instrument Performance |
| Effective Area |
0.23 cm2 peak @ 195Å for 170-210Å band
0.06 cm2 peak @ 270Å for 250-290Å band |
Expected line intensity from 1" × 1" area |
Quiet Sun |
8 photons/s for Fe XII 195Å
2 ph/s for He II 256Å |
| Active Region |
2 × 102 ph/s for Fe XII 195Å
1 × 101 ph/s for He II 256Å
1 × 102 ph/s for Fe XV 284Å |
| M2-class Flare |
4 × 104 ph/s for Fe XXIV 192Å
3 × 103 ph/s for Fe XXIV 255Å
4 × 103 ph/s for He II 256Å
1 × 103 ph/s for Fe XV 284Å |
| Accuracy of measurement |
Doppler velocity |
v = 2 km/s for 2 × 102 photons (Fe XV 284Å)
v = 3 km/s for 2 × 103photons (Fe XXIV 255Å) |
| Line width |
dw = 4 km/s for 2 × 102 photons (Fe XV 284Å)
dw = 9 km/s for 2 × 103 photons (Fe XXIV 255Å) |
| Data compression in MDP |
Bit compression: 16-to-12 bit or 14-to-12 bit compression |
| Image compression: DPCM (loss less) or 12bit-JPEG |
| Data rate |
1/8 of total telemetry (40kbps) after compression |
| Control of EIS observation |
Observing sequence in EIS ICU |
|